PT-2000 Operating Instructions

Operating Instructions

Probe Tack Tester

Model PT-2000

Setup, calibration, testing and maintenance instructions for the ChemInstruments PT-2000 Probe Tack Tester.

Revision 1.5, January 26, 2026. ChemInstruments, 510 Commercial Drive, Fairfield, Ohio 45014. (513) 860-1598

ChemInstruments PT-2000 Probe Tack Tester

Product Description

Congratulations on the purchase of the ChemInstruments Probe Tack Tester, a versatile, user-friendly instrument for determining probe tack values of adhesive laminates.

The unit has the following features:

  • Automated test sequence.
  • Collected test data can be exported via USB port.
  • Selectable units of measure: Kilograms, Grams, Newtons, Pounds, and Ounces.
  • Compatible with EZ-Data System software.

Specifications

Electrical85-265 VAC, 50/60 Hz, 2 amps
Operating Temperature32–122 °F (0–50 °C)
Humidity0–55% relative humidity
Speed2-30 in/min (1 in/min increments)
1-12 mm/sec (1 mm/sec increments)
Dwell Time0.5–30 seconds (0.5 second increments)
Dimensions (W x D x H)13 x 16 x 18 in (33 x 41 x 46 cm)
Weight25 lbs (12 kg)

Safety Warning

WARNING: This equipment can cause injury if not used properly. It is the operator's responsibility to read and follow all safety instructions.

ELECTRICAL WARNING: This is a dual-voltage machine. Ensure the unit is connected to the correct power supply (120 or 240 VAC). Incorrect power may cause equipment damage or electrical hazards.

Unpacking

Carefully unpack the Probe Tack Tester and inspect for shipping damage. If damage is found:

  • Take photographs
  • Retain all packaging
  • Notify the carrier immediately

Package Contents Checklist:

  • Test frame (with motor/drive and data acquisition system)
  • User manual (this document)
  • Probe and annular ring
  • Power cord

Installation Instructions

  1. Position the Test Frame. Place the Probe Tack Tester on a stable, level workbench. Ensure sufficient clearance for the unit's physical dimensions (refer to the Specifications section).
  2. Recommended Operating Environment. Maintain laboratory conditions of 72 ± 2 °F (22 ± 1 °C) and 50 ± 5% relative humidity (RH) for optimal performance and repeatability.
  3. Install the Annular Ring and Probe. Attach the probe to the threaded bolt on the load cell.
  4. Connect Power. Insert the power cord into the receptacle on the rear panel of the unit, then plug the opposite end into a properly grounded AC outlet (see electrical specifications).

Key Components

Power SwitchOn back panel, next to power input
USB PortFor data export and easy setup
Annular RingHolds test material
Load Cell AssemblyIncludes probe
Calibration HookUsed during calibration procedure
ProbeUsed to measure force of the test material
Touch Screen DisplayMain interface for control and results
Rear panel of the PT-2000 with the AC power input and switch and the USB port numbered 1 2
  1. 1AC Power Input & Switch
  2. 2USB
PT-2000 test head with the annular ring, probe, load cell assembly and calibration hook numbered 1 2 3 4
  1. 1Annular Ring
  2. 2Probe
  3. 3Load Cell Assembly
  4. 4Calibration Hook

Touch Screen Format

Main Screen

Button/LabelFunction
Current LoadDisplays current load cell reading
Current SpeedDisplays current test speed
Clean ProbeMoves probe to cleaning position
Start TestBegins test sequence
SetupAccess setup options
GraphShows graph and results from last test
StatisticsDisplays data from last test
PT-2000 Main screen with Current Load, Current Speed, Clean Probe, Setup, Graph and Statistics
Main screen, before the probe is cleaned.
PT-2000 Main screen with Current Load, Current Speed, Start Test, Setup, Graph and Statistics
Main screen, ready to start a test.

Setup Screen

ButtonFunctionLocked by D2979 Mode
Calibrate Load CellLoad cell calibrationNo
UnitsSet speed and force unitsNo
SpeedSet test speedYes
DwellSet dwell timeYes
ModeSelect test methodNo
Qualify OperationVerifies hardware functionNo
Clean ProbeMoves probe to cleaning positionNo
Calibrate Dwell TimeAllows the D 2979 dwell time tolerance band to be calibrated so that the dwell time falls within 0.99 and 1.01 secondsNo
AboutView software and hardware versionNo
ExitReturn to main screenNo

NOTE: After 30 seconds of inactivity in any setup screen (except Qualify Operation), the screen will return to the Main Screen.

PT-2000 Setup screen with Calibrate Load Cell, Units, Speed, Dwell, Mode, Qualify Operation, Clean Probe, Calibrate Dwell Time, About and Exit
Setup screen.

Graph Screen

Displays the following data from the last test:

  • Force vs. time graph
  • Minimum and maximum values

Statistics Screen

Shows test results, including:

  • Minimum and maximum
  • Variance and standard deviation
  • Work (area under the force-time curve)

Machine Setup

Power On the Instrument

WARNING: The instrument must operate within a temperature range of 32°F to 122°F (0°C to 50°C).

WARNING: Before applying power, ensure that the equipment is completely free of condensation, both inside and out. Moisture can cause internal damage or malfunction.

To power on the Probe Tack Tester, locate the master power switch on the back panel of the control cabinet, adjacent to the power line receptacle. Switch it to the ON position.

If the load cell assembly is not positioned at the home switch during startup, a warning message will appear. The system must establish the home position of the load cell assembly upon power-up to ensure proper operation.

PT-2000 Home System screen: the system needs to return to the HOME position, with a HOME button
Home system screen.

Load Cell Calibration

To ensure accurate and reliable test results, it is essential to calibrate the load cell before use. The Probe Tack Tester is equipped with an integrated calibration procedure, accessible through the touchscreen interface. Calibration should be performed:

  • During initial setup
  • After relocation or extended periods of inactivity
  • Whenever measurement accuracy is in question
  • As part of routine maintenance

IMPORTANT: Power on the Probe Tack Tester for at least 30 minutes prior to calibration to allow the system to stabilize.

NOTE: The calibration process uses grams as the default unit. Ensure calibration weights and inputs are in grams.

Calibration Procedure

Follow these steps to calibrate the load cell:

  1. From the Main Screen, tap Setup.
  2. Select Calibrate Load Cell.
  3. Review the description of the two-point calibration process and tap OK to continue.
  4. When prompted for the low calibration value (zero load), ensure no weight is hanging on the calibration hook, then tap OK.
  5. When prompted for the high calibration value, hang a certified calibration weight of 1000 grams on the calibration hook and make sure the weight is completely still before proceeding. Enter the exact weight in grams, tap Enter, then tap OK to confirm and continue.
  6. You will be returned to the Main Screen, where the calibrated reading will display under Current Load.
  7. To verify calibration, hang a different known weight on the calibration hook and check if the displayed load matches the expected value.
  8. If the reading is inaccurate, repeat the calibration process.
PT-2000 with a calibration weight hanging from the calibration hook under the load cell assembly
A calibration weight hanging from the calibration hook.

Units Setup

To adjust the units of force and speed:

  1. From the Main Screen, tap Setup.
  2. Select Units.
  3. Choose your desired units.
  4. Tap OK to save your selections.

Mode Setup

Selecting a specific test mode automatically configures the tester to meet the corresponding standard, locking the speed and dwell time.

  1. From the Main Screen, tap Setup.
  2. Select Mode.
  3. Choose appropriate method.
  4. Tap OK to confirm.
Test ModeSpeedDwell Time
D297924 in/min1 second
D297910 mm/sec1 second

Speed Setup

To set the load cell assembly travel speed for your test:

  1. From the Main Screen, tap Setup.
  2. Select Speed.
  3. Enter the desired value, and press Enter.
  4. Tap OK to confirm.

NOTE: Speed is fixed at 24 in/min or 10 mm/sec when ASTM D 2979 test mode is selected.

Dwell Time Setup

Dwell time refers to the duration the probe remains in contact with the test sample prior to separation.

  1. From the Main Screen, tap Setup.
  2. Select Dwell.
  3. Enter the desired value, and press Enter.
  4. Tap OK to confirm.

NOTE: Dwell time is fixed at 1 second when ASTM D 2979 test mode is selected.

Calibrate Dwell Time

This process allows the PT-2000 to calibrate the dwell time for the D2979 test method. The test method requires a dwell time of 1.00 second (± 0.01 seconds). After each test run, the PT-2000 calculates the actual dwell time. If the measured dwell time falls outside the acceptable range, the system automatically adjusts the timing for the next test cycle. Multiple test cycles may be required to bring the dwell time within specification.

PT-2000 Calibrate Dwell Time screen for D2979 mode, with Calibrate and Cancel buttons
Calibrate dwell time screen (D2979 mode).

Qualify Operation

The Qualify Operation screen is designed to verify the proper functionality of key system components.

Diagnostic Tools

  • Current Load: Displays the real-time force measured by the load cell in selected units.
  • Current Speed: Displays the current speed setting.
  • AD Reading: Shows raw analog-to-digital values from the load cell. Pulling down on the calibration hook gives values > 32000; pushing up on the calibration hook gives values < 32000.
  • Measured Dwell Time: Displays the measured dwell time from the previous test.
  • Test Speed: Runs a 4-inch travel test to verify load cell assembly speed. The system measures and reports the time taken to travel the set distance.
  • Verify Load Cell: Records load cell output for 10 seconds. No movement occurs during this test. The system displays a force graph and statistical summary to confirm load cell accuracy using a calibration weight.
  • Align Probe: Moves the probe to a position where its alignment can be checked.
PT-2000 Qualify Operation screen with Load Cell Reading, Measured Dwell Time, Verify Load Cell, Test Speed, Align Probe and Exit
Qualify operation screen.

Theory of Operation

A test sample is secured to the annular ring in accordance with the test method and placed on the test platform. The probe is then brought into contact with the test sample and retracted at a specified speed. An electronic load cell measures the applied force and transmits the data to the data acquisition unit.

The data acquisition unit records the force measurements and stores the data points in memory for calculation of maximum and minimum values. The collected data can be transferred to a PC through the USB connection using a compatible software program.

The load cell samples at a rate of 2,000 times per second. Two samples are averaged and stored as a single data point, resulting in one data point every millisecond. A maximum of 26,000 data points can be stored for each test.

Running a Test

Probe tack tests are conducted according to established test methods such as ASTM, PSTC, TLMI, and others. These tests are performed to determine the tack adhesive values of the selected pressure-sensitive adhesive (PSA) material as it is removed from a stainless-steel test probe.

NOTE: Ensure the load cell has been calibrated before conducting a test.

Cleaning the Probe

After powering on the PT-2000, the initial screen will prompt the user to perform a probe cleaning sequence. Press the CLEAN PROBE button to raise the probe to a position that allows cleaning using a cotton swab and the appropriate solvent.

After each test, the PT-2000 displays the test results on the graph screen. Pressing anywhere on the screen will initiate the CLEAN PROBE sequence so the probe can be cleaned before the next test.

During the cleaning sequence, the system senses when pressure is applied to the probe and starts a five-second timer, allowing the user to clean the probe for five seconds. Once the timer completes, the probe automatically lowers to the test position.

Test Procedure

The ChemInstruments Probe Tack Tester is simple to operate. Follow the steps below to perform a Probe Tack Test:

  1. Secure the test sample on the annular ring according to the applicable test method.
  2. Install the annular ring and press the START TEST button on the main screen. The probe will rise and contact the test material for one second. The probe will then retract while the load cell measures the test forces.
  3. When the test is complete, the graph screen will display the recorded high, low, and average values.
  4. Press anywhere on the graph screen to initiate the CLEAN PROBE sequence as described above.
  5. Repeat the procedure for additional tests.

Dwell Time Measurement

Dwell time is measured from the first data point that is less than −5 grams to the first data point that is greater than 0 grams. Data points are collected every one millisecond.

The PT-2000 automatically calculates the dwell time. Results can be viewed on the QUALIFY OPERATION screen under SETUP.

EZ Data Software System

EZ Data is a proprietary software application developed by ChemInstruments, designed to run on your computer and seamlessly interface with ChemInstruments testing equipment. This powerful tool enables users to:

  • Save test data files for archival or further analysis
  • Export raw test data directly to Microsoft Excel
  • Graph and crop data for clearer visual representation
  • Tabulate test data files for easy comparison
  • Overlay multiple test results to analyze performance trends

For detailed instructions on software features, setup, and operation, please consult the EZ Data User Manual.

The PT-2000 tester connects to your computer using a USB Type A-to-B cable, compliant with USB 2.0 (Revision 2.0) standards.

EZ Data Main Screen

The image below displays the main screen of the EZ Data software with a PT-2000 tester connected. This interface provides a comprehensive overview of the testing process, including:

  • Real-time load readings displayed as both a numeric value and a dynamic graph as data is received from the load cell
  • Current test setup parameters, clearly shown for reference

This intuitive layout ensures users can monitor, control, and modify tests with efficiency and precision.

EZ Data software main screen with a PT-2000 connected, showing current load, a force graph and the test setup
EZ Data main screen with a PT-2000 connected.

EZ Data Setup Screen

Test setup is streamlined within a single, centralized screen in EZ Data, providing a user-friendly interface for configuring all necessary test parameters. This unified setup screen allows users to efficiently define key settings such as test speed, dwell time, units, and mode, all without navigating through multiple menus. By consolidating setup functions into one location, EZ Data enhances workflow efficiency and reduces the potential for configuration errors, ensuring consistent and reliable test execution.

EZ Data PT-2000 Parameter Setup window with force units, speed units, speed, dwell time and mode
PT-2000 parameter setup in EZ Data.

Maintenance

Troubleshooting

The troubleshooting chart outlines a range of potential issues that may arise during regular use or over time. Each listed problem includes possible causes and recommended corrective actions to help quickly identify and resolve the issue. Once the specific problem has been determined, refer to the corresponding maintenance procedure to perform the necessary corrective steps. These procedures are designed to restore optimal performance and minimize equipment downtime.

ProblemPossible CauseProcedure
No data collectedNot in Run MenuRefer to running a test
Load Cell Assembly does not move during a testLoad cell assembly is in the upper idle positionMove the load cell assembly to the test position. Refer to test procedure
Motor is not allowing the assembly to moveReplace motor
Data measurement consistently low/highImproper calibrationCheck calibration
Bad calibrationRefer to load cell calibration
Calibration driftsBad or damaged load cellReplace load cell

Maintenance Procedures

As with all precision instruments, regular care and maintenance are essential to ensure accurate performance and extend the operational lifespan of the equipment. Consistent general cleaning and proper handling not only preserve the integrity of sensitive components but also help prevent malfunctions and minimize the risk of costly downtime. By adhering to recommended maintenance practices, users can maintain reliable test results and enjoy trouble-free operation over the long term.

Touchscreen Cleaning Guidelines

The touch panel is sensitive to chemicals and requires careful cleaning to maintain optimal functionality and appearance. Follow these specific cleaning instructions to ensure safe and effective care:

  • Use a soft, lint-free cloth for cleaning. The 3M Microfiber Lens Cleaning Cloth is highly recommended as it effectively cleans the touch panel without the need for liquid cleaners.
  • The cloth may be used dry or lightly dampened with a mild cleaner or ethanol. Ensure the cloth is only lightly damp, never wet.
  • Never apply cleaner directly to the touch panel surface. If any liquid accidentally contacts the panel, immediately blot it with an absorbent cloth to prevent seepage.
  • Use only neutral pH cleaners (neither acidic nor alkaline). Avoid any cleaners containing ammonia, phosphates, or ethylene glycol.
  • When applying cleaner, avoid contact with the edges of the film or glass and the flex cable tail to prevent damage.
  • Clean the surface gently, following the direction of any surface texture if present.
  • Do not use harsh chemicals such as paint thinner, acetone, toluene, xylene, kerosene, or strong alcohols like propyl or isopropyl alcohol in high concentrations.
  • Suitable cleaning products are commercially available and specifically formulated for delicate touch panels. Examples include Klear Screen™ and retail brands like Glass Plus® Glass and Surface Cleaner by Reckitt Benckiser.
  • Most commercial products contain 1–3% isopropyl alcohol by volume, which is generally safe for use on resistive touch panels.

IMPORTANT: Always check the product's ingredient list carefully. Many cleaners contain unacceptable substances such as ammonia, phosphates, or ethylene glycol, which can damage the touch panel's optical clarity and functionality.

Load Cell Assembly Removal for Recertification

The load cell assembly can be removed from the tester for recertification or servicing. To ensure safe and proper removal, follow these steps carefully:

  1. Prepare the work area: Make sure the equipment is powered off and disconnected from any power sources before beginning the removal process.
  2. Support the load cell assembly: Before loosening any fasteners, place one hand firmly under the load cell assembly to support its weight and prevent accidental drops.
  3. Remove the mounting screws: Using the appropriate tool, carefully remove the two large screws securing the load cell assembly. Maintain your support on the assembly as you loosen the screws to avoid sudden movement or damage.
  4. Gently disengage the assembly: Once the screws are fully removed, slowly pull the load cell assembly backward to expose the electrical connector.
  5. Disconnect the connector: Carefully unplug the connector by holding the connector housing itself. Do not pull on the wires to avoid damaging the cable or connection points.
  6. Inspect and package: After removal, inspect the assembly for any visible damage before packaging it securely for shipment. Use appropriate protective materials to prevent impact or static damage during transport.

By following these steps, you help ensure the integrity of the load cell assembly and maintain its accuracy after recertification.

PT-2000 load cell assembly pulled back from the mast to expose the wiring connector
The load cell connector, exposed once the assembly is pulled back.

Replacement Parts

Part NumberDescription
R-EZ DATA 5 BOARDReplacement board for 2000 series machines
R-POWER SUPPLY (65)Replacement power supply 65W
PT-2000 LOAD CELL 2LBReplacement PT-2000 load cell 2 lb
PT-2000 LOAD CELL 5LBReplacement PT-2000 load cell 5 lb
PT-2000 LOAD CELL 10LBReplacement PT-2000 load cell 10 lb
PT-2000 LOAD CELL 20LBReplacement PT-2000 load cell 20 lb
R-PT ANNULAR RINGReplacement annular ring
R-PT-2000 PROBEReplacement PT-2000 probe

Ready for a quote?

Tell us what you test and we will quote the probe tack tester, load cells, probes, annular rings, calibration, service or replacement parts you need.

Request a Quote