LT-2000 Operating Instructions

Operating Instructions

Loop Tack Tester

Model LT-2000

Setup, calibration, testing and maintenance instructions for the ChemInstruments LT-2000 Loop Tack Tester.

Revision 1.4, May 23, 2025. ChemInstruments, 510 Commercial Drive, Fairfield, Ohio 45014. (513) 860-1598

ChemInstruments LT-2000 Loop Tack Tester

Product Description

Congratulations on the purchase of the ChemInstruments Loop Tack Tester, a versatile, user-friendly instrument for determining loop tack values of adhesive laminates.

The unit has the following features:

  • Automated test sequence.
  • Collected test data can be exported via USB port.
  • Selectable units of measure: Kilograms, Grams, Newtons, Pounds, and Ounces.
  • Compatible with EZ-Data System software.

Specifications

Electrical85-265 VAC, 50/60 Hz, 2 amps
Operating Temperature32–122 °F (0–50 °C)
Humidity0–55% relative humidity
Speed2-30 in/min (1 in/min increments)
60-720 mm/min (1 mm/min increments)
Dwell Time0–30 seconds (0.5 second increments)
Loop Length5 inch or 7 inch
Test MethodsASTM D6195, FTM9, or Custom
Dimensions (W x D x H)13 x 16 x 18 in (33 x 41 x 46 cm)
Weight25 lbs (12 kg)

Safety Warning

WARNING: This equipment can cause injury if not used properly. It is the operator's responsibility to read and follow all safety instructions.

ELECTRICAL WARNING: This is a dual-voltage machine. Ensure the unit is connected to the correct power supply (120 or 240 VAC). Incorrect power may cause equipment damage or electrical hazards.

Unpacking

Carefully unpack the Loop Tack Tester and inspect for shipping damage. If damage is found:

  • Take photographs
  • Retain all packaging
  • Notify the carrier immediately

Package Contents Checklist:

  • Test frame (with motor/drive and data acquisition system)
  • User manual (this document)
  • Grip and fastening pin
  • Power cord

Installation Instructions

  1. Position the Test Frame. Place the Loop Tack Tester on a stable, level workbench. Ensure sufficient clearance for the unit's physical dimensions (refer to the Specifications section).
  2. Recommended Operating Environment. Maintain laboratory conditions of 72 ± 2 °F (22 ± 1 °C) and 50 ± 5% relative humidity (RH) for optimal performance and repeatability.
  3. Install the Grip. Attach the grip to the load cell assembly using the supplied mounting pin. Ensure the knob is facing forward for correct orientation.
  4. Connect Power. Insert the power cord into the receptacle on the rear panel of the unit, then plug the opposite end into a properly grounded AC outlet (see electrical specifications).
LT-2000 load cell assembly with the mounting pin and grip numbered 1 2
  1. 1Mounting Pin
  2. 2Grip

Key Components

Power SwitchOn back panel, next to power input
USB PortFor data export and easy setup
Test Panel BracketHolds a 1 inch wide test panel
Load Cell AssemblyIncludes mounting bracket and grip
GripSecures the test loop to the load cell
MastTrack for the load cell during testing
Touch Screen DisplayMain interface for control and results
Rear panel of the LT-2000 with the AC power input and switch and the USB port numbered 1 2
  1. 1AC Power Input & Switch
  2. 2USB
Front of the LT-2000 with the mast, load cell assembly, grip and test panel bracket numbered 1 2 3 4
  1. 1Mast
  2. 2Load Cell Assembly
  3. 3Grip
  4. 4Test Panel Bracket

Touch Screen Format

Main Screen

Button/LabelFunction
Current LoadDisplays current load cell reading
Current SpeedDisplays current test speed
Start TestBegins test sequence
SetupAccess setup options
GraphShows graph and results from last test
StatisticsDisplays data from last test
LT-2000 touch screen Main screen with Current Load, Current Speed, Start Test, Setup, Graph and Statistics
Main screen.

Setup Screen

ButtonFunctionLocked by D6195 Mode
Calibrate Load CellLoad cell calibrationNo
UnitsSet speed and force unitsNo
SpeedSet test speedYes
DwellSet dwell timeYes
ModeSelect test methodNo
Qualify OperationVerifies hardware functionNo
AboutView software and hardware versionNo
ExitReturn to main screenNo

NOTE: After 30 seconds of inactivity in any setup screen (except Qualify Operation), the screen will return to the Main Screen.

LT-2000 Setup screen with Calibrate Load Cell, Units, Speed, Dwell, Mode, Qualify Operation, About and Exit
Setup screen.

Graph Screen

Displays the following data from the last test:

  • Force vs. time graph
  • Minimum and maximum values

Statistics Screen

Shows test results, including:

  • Minimum and maximum
  • Variance and standard deviation
  • Work (area under the force-time curve)

Machine Setup

Power On the Instrument

WARNING: The instrument must operate within a temperature range of 32°F to 122°F (0°C to 50°C).

WARNING: Before applying power, ensure that the equipment is completely free of condensation, both inside and out. Moisture can cause internal damage or malfunction.

To power on the Loop Tack Tester, locate the master power switch on the back panel of the control cabinet, adjacent to the power line receptacle. Switch it to the ON position.

If the load cell assembly is not positioned at the home switch during startup, a warning message will appear. The system must establish the home position of the load cell assembly upon power-up to ensure proper operation.

LT-2000 Home System screen: the system needs to return to the HOME position, with a HOME button
Home system screen.

Load Cell Calibration

To ensure accurate and reliable test results, it is essential to calibrate the load cell before use. The Loop Tack Tester is equipped with an integrated calibration procedure, accessible through the touchscreen interface. Calibration should be performed:

  • During initial setup
  • After relocation or extended periods of inactivity
  • Whenever measurement accuracy is in question
  • As part of routine maintenance

IMPORTANT: Power on the Loop Tack Tester for at least 30 minutes prior to calibration to allow the system to stabilize.

NOTE: The calibration process uses grams as the default unit. Ensure calibration weights and inputs are in grams.

Calibration Procedure

Follow these steps to calibrate the load cell:

  1. Ensure the grip is securely attached to the load cell using the mounting pin.
  2. From the Main Screen, tap Setup.
  3. Select Calibrate Load Cell.
  4. Review the description of the two-point calibration process and tap OK to continue.
  5. When prompted for the low calibration value (zero load), ensure no weight is placed on the calibration pad, then tap OK.
  6. When prompted for the high calibration value, place a certified calibration weight of 1000 grams on the calibration pad. Enter the exact weight in grams, tap Enter, then tap OK to confirm and continue.
  7. You will be returned to the Main Screen, where the calibrated reading will display under Current Load.
  8. To verify calibration, place a different known weight on the calibration pad and check if the displayed load matches the expected value.
  9. If the reading is inaccurate, repeat the calibration process.
LT-2000 with a calibration weight resting on the calibration pad of the load cell assembly 1
  1. 1Calibration Pad, shown with a calibration weight resting on it

Units Setup

To adjust the units of force and speed:

  1. From the Main Screen, tap Setup.
  2. Select Units.
  3. Choose your desired units.
  4. Tap OK to save your selections.

NOTE: Speed units are fixed at mm/min when FTM9 test mode is selected.

Mode Setup

Selecting a specific test mode automatically configures the tester to meet the corresponding standard, locking the speed and dwell time.

  1. From the Main Screen, tap Setup.
  2. Select Mode.
  3. Choose appropriate method.
  4. Tap OK to confirm.
Test ModeSpeedDwell TimeLoop Length
D6195 – 5″12 in/min05″
D6195 – 5″300 mm/min05″
Custom – 5″selectableselectable5″
D6195 – 7″12 in/min07″
D6195 – 7″300 mm/min07″
FTM9300 mm/min07″
Custom – 7″selectableselectable7″

Speed Setup

To set the load cell assembly travel speed for your test:

  1. From the Main Screen, tap Setup.
  2. Select Speed.
  3. Enter the desired value, and press Enter.
  4. Tap OK to confirm.

NOTE: Speed is fixed at 12 in/min or 300 mm/min when ASTM D 6195 or FTM9 test mode is selected.

Dwell Time Setup

Dwell time refers to the duration the test loop remains in contact with the panel prior to separation.

  1. From the Main Screen, tap Setup.
  2. Select Dwell.
  3. Enter the desired value, and press Enter.
  4. Tap OK to confirm.

NOTE: Dwell time is fixed at 0 seconds when ASTM D 6195 or FTM9 test mode is selected.

Grip Operation

To operate the grip:

  • Open the grip by turning the thumb screw counter-clockwise.
  • Insert the free end of the looped specimen into the grip.
  • Close the grip by turning the thumb screw clockwise until secure.

IMPORTANT: The grip connects to the Load Cell Assembly via a mounting pin and yoke. Always remove the grip from the assembly to load test specimens safely.

Diagram of the grip open and closed, showing the thumb screw and washer
Grip open (left) and grip closed (right).

WARNING: Avoid rotating the yoke on the threaded mounting rod. If the yoke contacts the wall of the load cell housing, permanent damage may occur. Maintain a visible gap between the yoke and the housing to allow the load cell to operate correctly.

Qualify Operation

The Qualify Operation screen is designed to verify the proper functionality of key system components.

Diagnostic Tools

  • Current Load: Displays the real-time force measured by the load cell in selected units.
  • Current Speed: Displays the current speed setting.
  • AD Reading: Shows raw analog-to-digital values from the load cell. Pulling down on the grip gives values > 32000; pushing up on the grip gives values < 32000.
  • Test Speed: Runs a 4-inch travel test to verify load cell assembly speed. The system measures and reports the time taken to travel the set distance.
  • Verify Load Cell: Records load cell output for 10 seconds. No movement occurs during this test. The system displays a force graph and statistical summary to confirm load cell accuracy using a calibration weight.

Theory of Operation

The ChemInstruments Loop Tack Tester measures the tack force of pressure-sensitive adhesive (PSA) materials using a standardized loop tack test method. The procedure is based on the principles outlined in ASTM D6195 – Standard Test Methods for Loop Tack.

A PSA sample is formed into a loop with the adhesive surface facing outward. This loop is brought into contact with a clean stainless steel test panel, then immediately retracted at a fixed speed. The instrument's load cell records the peak force required to separate the loop from the panel, an indication of the material's initial tack characteristics.

  • Testing adheres to ASTM D6195, which specifies procedures and apparatus requirements to ensure accurate, repeatable results.
  • The LT-2000 is programmed with a dedicated ASTM D6195 mode, automatically setting the speed and dwell time according to the standard (typically 12 inches/min and 0-second dwell).
  • The load cell samples force data at 400 Hz, with every 8 samples averaged into one data point collected every 20 milliseconds.
  • Up to 13,500 data points can be stored per test and exported via USB for analysis.

Test Procedure

Loop tack tests should follow the guidelines set forth in ASTM D6195, which provides standardized methods for evaluating the tack of PSA materials. The ChemInstruments LT-2000 includes a built-in ASTM D6195 test mode to ensure compliance with these parameters.

NOTE: ASTM D6195 specifies a standard test speed of 12 inches per minute (300 mm/min) and a 0-second dwell time. These settings are automatically applied when ASTM D6195 mode is selected in the setup menu and cannot be altered.

Follow the steps below to conduct a loop tack test according to ASTM D6195:

  1. Clean the test panel using a solvent such as Methyl Alcohol or Methyl Ethyl Ketone. Ensure the panel is 304 Bright Annealed Stainless Steel with a surface roughness of 2 µin ± 1 µin, as specified by the standard.
  2. Insert the cleaned panel into the test panel holder.
  3. Remove the grip from the load cell assembly.
  4. Prepare a test specimen: Cut a 1 in × 5 in strip (25 mm × 127 mm) with the long edge in the machine direction.
  5. Remove the release liner completely.
  6. Form a loop by gently folding the sample back onto itself without creasing, adhesive side out, forming a teardrop shape.
  7. Secure the ends using ½ inch wide masking tape.
  8. Insert the loop into the grip so it touches the spring block, ensuring proper alignment.
  9. Reinstall the grip into the mounting yoke using the mounting pin.
  10. On the touchscreen, select ASTM D6195 mode in the Setup menu (if not already selected), then press START TEST.

The LT-2000 will automatically execute the test according to ASTM D6195 parameters and display the results.

TIP: After each test, slide the test panel to expose a fresh, unused surface. Avoid overlapping previous test spots to ensure test accuracy, as outlined in best practices derived from ASTM guidance.

Test Results and Data Handling

Upon completion of a test, the system automatically calculates and displays the results in both graphical and text formats. This dual-format presentation allows users to quickly analyze performance characteristics and evaluate individual test outcomes with clarity.

For Loop Tack Tests, please note that the average test value is not calculated due to the nature of the test methodology, which is designed to capture peak force values rather than averaged data.

If the EZ Data software is connected to the system, all test data will be automatically uploaded for digital storage. This includes raw data, test parameters, and results, which can be reviewed, exported, or archived for quality control, reporting, or traceability purposes.

Alternate Test Modes

The Loop Tack Tester is equipped with a variety of test modes to accommodate diverse testing standards and user requirements. Available modes include:

  • ASTM D6195 – 5-inch loop
  • ASTM D6195 – 7-inch loop
  • FTM 9 (Federal Test Method 147) – 7-inch loop
  • Custom 5-inch loop
  • Custom 7-inch loop

The custom loop test modes offer enhanced flexibility, allowing users to define specific test parameters such as test speed and dwell time, ensuring the tester can be configured to meet unique or specialized application needs. This adaptability makes the Loop Tack Tester suitable for both standardized compliance testing and research or product development scenarios.

Test Start Positions

Selecting a test mode automatically sets the starting position of the load cell assembly to optimize test performance and accuracy based on loop length.

7-inch loop tests

The load cell assembly is positioned at the uppermost location to accommodate the full travel required for the longer loop. This ensures that the entire range of motion is available for testing and data collection.

5-inch loop tests

The load cell assembly is positioned lower to shorten the travel distance. This reduces unnecessary movement, minimizes cycle time, and enhances test efficiency for the shorter loop length.

These start positions are maintained until a new loop length mode is selected. Always confirm that the load cell resets to the correct home position after changing modes to prevent measurement errors.

Overload Monitoring

The Loop Tack Tester is equipped with comprehensive overload monitoring to protect critical system components and ensure operational safety.

During downward movement of the load cell assembly, the system continuously monitors for negative force overloads, helping to prevent potential damage to the load cell and lead screw. Additionally, while a test is actively running, the system evaluates the applied force in real time. If an excessive force is detected, the test is immediately halted to prevent overload-related damage.

In either case, the system will:

  • Instantly stop all motion
  • Display an error message
  • Require the user to HOME the system before resuming operation

For test-related overloads, all collected data up to the point of the interruption is preserved. This data can be reviewed through the Graph or Statistics screens, allowing users to assess the conditions that led to the overload and make informed adjustments for future tests.

EZ Data Software System

EZ Data is a proprietary software application developed by ChemInstruments, designed to run on your computer and seamlessly interface with ChemInstruments testing equipment. This powerful tool enables users to:

  • Save test data files for archival or further analysis
  • Export raw test data directly to Microsoft Excel
  • Graph and crop data for clearer visual representation
  • Tabulate test data files for easy comparison
  • Overlay multiple test results to analyze performance trends

For detailed instructions on software features, setup, and operation, please consult the EZ Data User Manual.

The LT-2000 tester connects to your computer using a USB Type A-to-B cable, compliant with USB 2.0 (Revision 2.0) standards.

EZ Data Main Screen

The image below displays the main screen of the EZ Data software with an LT-2000 tester connected. This interface provides a comprehensive overview of the testing process, including:

  • Real-time load readings displayed as both a numeric value and a dynamic graph as data is received from the load cell
  • Current test setup parameters, clearly shown for reference

This intuitive layout ensures users can monitor, control, and modify tests with efficiency and precision.

EZ Data software main screen with an LT-2000 connected, showing current load, a force graph and the test setup
EZ Data main screen with an LT-2000 connected.

EZ Data Setup Screen

Test setup is streamlined within a single, centralized screen in EZ Data, providing a user-friendly interface for configuring all necessary test parameters. This unified setup screen allows users to efficiently define key settings such as test speed, dwell time, units, and mode, all without navigating through multiple menus. By consolidating setup functions into one location, EZ Data enhances workflow efficiency and reduces the potential for configuration errors, ensuring consistent and reliable test execution.

EZ Data LT-2000 Parameter Setup window with force units, speed units, speed, dwell time and mode
LT-2000 parameter setup in EZ Data.

Maintenance

Troubleshooting

The troubleshooting chart outlines a range of potential issues that may arise during regular use or over time. Each listed problem includes possible causes and recommended corrective actions to help quickly identify and resolve the issue. Once the specific problem has been determined, refer to the corresponding maintenance procedure to perform the necessary corrective steps. These procedures are designed to restore optimal performance and minimize equipment downtime.

ProblemPossible CauseCorrective Action
Load Cell Assembly does not move during a testMotor is not allowing the assembly to moveReplace motor
Data measurement consistently low/highImproper calibrationCheck calibration
Bad calibrationRefer to load cell calibration
Calibration driftsBad or damaged load cellReplace load cell

Maintenance Procedures

As with all precision instruments, regular care and maintenance are essential to ensure accurate performance and extend the operational lifespan of the equipment. Consistent general cleaning and proper handling not only preserve the integrity of sensitive components but also help prevent malfunctions and minimize the risk of costly downtime. By adhering to recommended maintenance practices, users can maintain reliable test results and enjoy trouble-free operation over the long term.

Touchscreen Cleaning Guidelines

The touch panel is sensitive to chemicals and requires careful cleaning to maintain optimal functionality and appearance. Follow these specific cleaning instructions to ensure safe and effective care:

  • Use a soft, lint-free cloth for cleaning. The 3M Microfiber Lens Cleaning Cloth is highly recommended as it effectively cleans the touch panel without the need for liquid cleaners.
  • The cloth may be used dry or lightly dampened with a mild cleaner or ethanol. Ensure the cloth is only lightly damp, never wet.
  • Never apply cleaner directly to the touch panel surface. If any liquid accidentally contacts the panel, immediately blot it with an absorbent cloth to prevent seepage.
  • Use only neutral pH cleaners (neither acidic nor alkaline). Avoid any cleaners containing ammonia, phosphates, or ethylene glycol.
  • When applying cleaner, avoid contact with the edges of the film or glass and the flex cable tail to prevent damage.
  • Clean the surface gently, following the direction of any surface texture if present.
  • Do not use harsh chemicals such as paint thinner, acetone, toluene, xylene, kerosene, or strong alcohols like propyl or isopropyl alcohol in high concentrations.
  • Suitable cleaning products are commercially available and specifically formulated for delicate touch panels. Examples include Klear Screen™ and retail brands like Glass Plus® Glass and Surface Cleaner by Reckitt Benckiser.
  • Most commercial products contain 1–3% isopropyl alcohol by volume, which is generally safe for use on resistive touch panels.

IMPORTANT: Always check the product's ingredient list carefully. Many cleaners contain unacceptable substances such as ammonia, phosphates, or ethylene glycol, which can damage the touch panel's optical clarity and functionality.

Load Cell Assembly Removal for Recertification

The load cell assembly can be removed from the tester for recertification or servicing. To ensure safe and proper removal, follow these steps carefully:

  1. Prepare the work area: Make sure the equipment is powered off and disconnected from any power sources before beginning the removal process.
  2. Support the load cell assembly: Before loosening any fasteners, place one hand firmly under the load cell assembly to support its weight and prevent accidental drops.
  3. Remove the mounting screws: Using the appropriate tool, carefully remove the two large screws securing the load cell assembly. Maintain your support on the assembly as you loosen the screws to avoid sudden movement or damage.
  4. Gently disengage the assembly: Once the screws are fully removed, slowly pull the load cell assembly backward to expose the electrical connector.
  5. Disconnect the connector: Carefully unplug the connector by holding the connector housing itself. Do not pull on the wires to avoid damaging the cable or connection points.
  6. Inspect and package: After removal, inspect the assembly for any visible damage before packaging it securely for shipment. Use appropriate protective materials to prevent impact or static damage during transport.

By following these steps, you help ensure the integrity of the load cell assembly and maintain its accuracy after recertification.

LT-2000 load cell assembly pulled back from the mast to expose the wiring connector
The load cell connector, exposed once the assembly is pulled back.

Replacement Parts

Part NumberDescription
R-EZ DATA 5 BOARDReplacement board for 2000 series machines
R-POWER SUPPLY (65)Replacement power supply 65W
LT-2000 LOAD CELL 2LBReplacement LT-2000 load cell 2 lb, no jaws
LT-2000 LOAD CELL 5LBReplacement LT-2000 load cell 5 lb, no jaws
LT-2000 LOAD CELL 10LBReplacement LT-2000 load cell 10 lb, no jaws
LT-2000 LOAD CELL 20LBReplacement LT-2000 load cell 20 lb, no jaws