Operating Instructions
Coefficient of Friction Tester
Model COF-2000
Setup, calibration, testing and maintenance instructions for the ChemInstruments COF-2000 Coefficient of Friction Tester.

Contents
Product Description
Congratulations on the purchase of the ChemInstruments Coefficient of Friction Tester, a versatile, user-friendly instrument for determining coefficient of friction values of many different materials according to many test methods including ASTM D-1894.
The unit has the following features:
- One step test sequence.
- Collected test data can be exported via USB port.
- Selectable units of measure: Kilograms, Grams, Newtons, Pounds, and Ounces.
- Compatible with EZ-Data System software.
Specifications
| Electrical | 85-265 VAC, 50/60 Hz, 2 amps |
|---|---|
| Operating Temperature | 32–122 °F (0–50 °C) |
| Humidity | 0–55% relative humidity |
| Speed | 1-24 in/min (1 in/min increments) 3-60 cm/min (1 cm/min increments) |
| Sled Mass | 1 – 2000 grams, 1 gram increments |
| Static Length | 1 – 10 seconds, 1 second increments |
| Test Methods | ASTM D-1894 |
| Dimensions (W x D x H) | 24 x 12 x 8 in (58 x 30 x 20 cm) |
| Weight | 29 lbs (13 kg) |
Safety Warning
WARNING: This equipment can cause injury if not used properly. It is the operator's responsibility to read and follow all safety instructions.
ELECTRICAL WARNING: This is a dual-voltage machine. Ensure the unit is connected to the correct power supply (120 or 240 VAC). Incorrect power may cause equipment damage or electrical hazards.
Unpacking
Carefully unpack the Coefficient of Friction Tester and inspect for shipping damage. If damage is found:
- Take photographs
- Retain all packaging
- Notify the carrier immediately
Package Contents Checklist:
- Test frame (with motor/drive and data acquisition system)
- User manual (this document)
- Test Sled with connecting cable
- Calibration cable
- Power cord
Installation Instructions
- Position the Test Frame. Place the Coefficient of Friction Tester on a stable, level workbench. Ensure sufficient clearance for the unit's physical dimensions (refer to the Specifications section).
- Recommended Operating Environment. Maintain laboratory conditions of 72 ± 2 °F (22 ± 1 °C) and 50 ± 5% relative humidity (RH) for optimal performance and repeatability.
- Level the COF-2000. Adjust the four leveling feet located at each corner of the machine.
- Connect Power. Insert the power cord into the receptacle on the rear panel of the unit, then plug the opposite end into a properly grounded AC outlet (see electrical specifications).
Key Components

| Power Switch | On back panel, next to power input |
|---|---|
| USB Port | For data export and easy setup |
| Test Sled | A 200-gram block wrapped with foam rubber that slides across the test platform during COF testing |
| Load Cell Assembly | Measures the friction forces during a COF test |
| Calibration Cable & Pulley | Used during calibration |
| E-Stop | Emergency stop button |
| Touch Screen Display | Main interface for control and results |
Touch Screen Format
Main Screen
| Button/Label | Function |
|---|---|
| Current Load | Displays current load cell reading |
| Current Speed | Displays current test speed |
| Sled Movement | Access sled movement options |
| Setup | Access setup options |
| Graph | Shows graph and results from last test |
| Statistics | Displays data from last test |

Sled Movement Screen
| Button/Label | Function |
|---|---|
| Current Load | Displays current load cell reading |
| Current Speed | Displays current test speed |
| Start Test | Begin test sequence |
| Return | Return to previous test's data collection location |
| Jog Left | Move the cart left at 15 IPM |
| Main | Return to main screen |
| Jog Right | Move the cart right at 15 IPM |

Setup Screen
| Button | Function |
|---|---|
| Calibrate Load Cell | Load cell calibration |
| Units | Set speed and force units |
| Speed | Set test platform travel speed |
| Sled Mass | Set test sled mass in grams |
| Static Length | Set length of static test in seconds |
| Qualify Operation | Verifies hardware function |
| About | View software and hardware version |
| Exit | Return to main screen |
NOTE: After 30 seconds of inactivity in any setup screen (except Qualify Operation), the screen will return to the Main Screen.

Graph Screen
Displays the following data from the last test:
- Force vs. time graph
- Minimum and maximum values
Statistics Screen
Shows test results, including:
- Minimum and maximum
- Variance and standard deviation
- Work (area under the force-time curve)
- First Peak
- Static COF
- Kinetic COF


Machine Setup
Power On the Instrument
WARNING: The instrument must operate within a temperature range of 32°F to 122°F (0°C to 50°C).
WARNING: Before applying power, ensure that the equipment is completely free of condensation, both inside and out. Moisture can cause internal damage or malfunction.
To power on the Coefficient of Friction Tester, locate the master power switch on the back panel of the control cabinet, adjacent to the power line receptacle. Switch it to the ON position.
If the test platform is not positioned at the home switch during startup, a warning message will appear. The system must establish the home position of the test platform upon power-up to ensure proper operation.

Load Cell Calibration
To ensure accurate and reliable test results, it is essential to calibrate the load cell before use. The Coefficient of Friction Tester is equipped with an integrated calibration procedure, accessible through the touchscreen interface. Calibration should be performed:
- During initial setup
- After relocation or extended periods of inactivity
- Whenever measurement accuracy is in question
- As part of routine maintenance
IMPORTANT: Power on the Coefficient of Friction Tester for at least 30 minutes prior to calibration to allow the system to stabilize.
NOTE: The calibration process uses grams as the default unit. Ensure calibration weights and inputs are in grams.
Calibration Procedure
Follow these steps to calibrate the load cell:
- Locate the wire attached to the load cell assembly.
- From the Main Screen, tap Setup.
- Select Calibrate Load Cell.
- Review the description of the two-point calibration process and tap OK to continue.
- When prompted for the low calibration value (zero load), ensure no weight is hanging on the calibration wire, then tap OK.
- When prompted for the high calibration value, hang a certified calibration weight of 1000 grams and verify that it is completely at rest. Enter the exact weight in grams, tap Enter, then tap OK to confirm and continue.
- You will be returned to the Main Screen, where the calibrated reading will display under Current Load.
- To verify calibration, hang a different known weight on the calibration wire and check if the displayed load matches the expected value.
- If the reading is inaccurate, repeat the calibration process.
Units Setup
To adjust the units of force and speed:
- From the Main Screen, tap Setup.
- Select Units.
- Choose your desired units.
- Tap OK to save your selections.
Speed Setup
To set the test platform travel speed for your test:
- From the Main Screen, tap Setup.
- Select Speed.
- Enter the desired value, and press Enter.
- Tap OK to confirm.
Sled Mass Setup
To set the mass of the test sled:
- From the Main Screen, tap Setup.
- Select Sled Mass.
- Enter the desired value, and press Enter.
- Tap OK to confirm.
Static Length Setup
The static length is variable between 1 and 10 seconds. This defines the amount of data collected for the static portion of the test.
To set the static length:
- From the Main Screen, tap Setup.
- Select Static Length.
- Enter the desired value, and press Enter.
- Tap OK to confirm.
Qualify Operation
The Qualify Operation screen is designed to verify the proper functionality of key system components.
Diagnostic Tools
- Current Load: Displays the real-time force measured by the load cell in selected units.
- Current Speed: Displays the current speed setting.
- Load Cell Reading: Shows raw analog-to-digital values from the load cell. Pulling to the right on the calibration hook gives values > 32000; pushing to the left on the calibration hook gives values < 32000.
- Test Speed: Runs a 4-inch travel test to verify test platform speed. The system measures and reports the time taken to travel the set distance.
- Verify Load Cell: Records load cell output for 10 seconds. No movement occurs during this test. The system displays a force graph and statistical summary to confirm load cell accuracy using a calibration weight.

Theory of Operation
During a COF test, the Test Platform moves at a constant speed while the Test Sled remains stationary, creating friction between the two surfaces. An electronic load cell measures the force and sends the data to the data acquisition unit.
The data acquisition unit collects and stores the force measurements in memory. This data is used to calculate the maximum, minimum, and average force values, as well as the Static COF and Kinetic COF of the test material.
The load cell samples at a rate of 400 times per second. Eight samples are averaged and stored as one data point, resulting in a data point every 20 milliseconds.
A maximum of 10,000 data points can be saved for each test.
Running a Test
Ensure the load cell has been calibrated and is reading correctly before performing a test.
Test Procedure
The following procedure describes a standard test sequence for the COF-2000 Tester and is typically performed in accordance with applicable industry standards, such as ASTM D1894, Standard Test Method for Static and Kinetic Coefficients of Friction of Plastic Film and Sheeting (or other relevant test methods as required).
- Attach the test sample to both the Test Sled and the Test Platform in accordance with the applicable test method.
- Place the Test Sled in the center of the Test Platform and as far to the left as the connecting cable allows.
- Align the Test Sled so that its edges are parallel to the edges of the Test Platform.
- Adjust the Test Sled position to provide sufficient slack in the connecting cable so it rests on the Test Platform. The Instant Force reading should be below 5 grams at this point.
WARNING: Before proceeding, ensure there are no objects in the path of the moving Test Platform.
- Press the START TEST button to begin the test sequence.
- Allow the Test Platform to travel and come to a complete stop at the end of the test.
- The calculated Static COF and Kinetic COF values will be displayed on the STATISTICS screen.
WARNING: Remove the Test Sled from the path of the Test Platform before pressing the RETURN button.
- Press the RETURN button to move the Test Platform back to the starting position for the next test.
COF Test Description and Results
During a COF test, the Test Platform moves at a constant speed while the Test Sled remains stationary. This movement creates friction between the two surfaces. The force produced by this friction is measured by the load cell and used to calculate the Static COF and Kinetic COF of the test material.
Static COF
The static portion of the test occurs first. The static length is set in the setup screen and can range from 1 to 10 seconds. The highest force measured during this time is saved as the First Peak. The Static COF is calculated using the formula:
Static COF = First Peak (in grams) ÷ Sled Mass (in grams)
Kinetic COF
The total test length is 7 inches. The kinetic portion of the test begins after the static portion is complete. The number of data points collected during the kinetic portion depends on the test speed. The average, high, and low force values are calculated from the kinetic data and displayed on the EZ Lab graph screen. The Kinetic COF is calculated as follows:
Kinetic COF = Average (in grams) ÷ Sled Mass (in grams)
EZ Data Software System
EZ Data is a proprietary software application developed by ChemInstruments, designed to run on your computer and seamlessly interface with ChemInstruments testing equipment. This powerful tool enables users to:
- Save test data files for archival or further analysis
- Export raw test data directly to Microsoft Excel
- Graph and crop data for clearer visual representation
- Tabulate test data files for easy comparison
- Overlay multiple test results to analyze performance trends
For detailed instructions on software features, setup, and operation, please consult the EZ Data User Manual.
The COF-2000 tester connects to your computer using a USB Type A-to-B cable, compliant with USB 2.0 (Revision 2.0) standards.
EZ Data Main Screen
The image below displays the main screen of the EZ Data software with a COF-2000 tester connected. This interface provides a comprehensive overview of the testing process, including:
- Real-time load readings displayed as both a numeric value and a dynamic graph as data is received from the load cell
- Current test setup parameters, clearly shown for reference
This intuitive layout ensures users can monitor, control, and modify tests with efficiency and precision.

EZ Data Setup Screen
Test setup is streamlined within a single, centralized screen in EZ Data, providing a user-friendly interface for configuring all necessary test parameters. This unified setup screen allows users to efficiently define key settings such as test speed, dwell time, units, and mode, all without navigating through multiple menus. By consolidating setup functions into one location, EZ Data enhances workflow efficiency and reduces the potential for configuration errors, ensuring consistent and reliable test execution.

Maintenance
Troubleshooting
The troubleshooting chart outlines a range of potential issues that may arise during regular use or over time. Each listed problem includes possible causes and recommended corrective actions to help quickly identify and resolve the issue. Once the specific problem has been determined, refer to the corresponding maintenance procedure to perform the necessary corrective steps. These procedures are designed to restore optimal performance and minimize equipment downtime.
| Problem | Possible Cause | Procedure |
|---|---|---|
| No data collected | Display is in SETUP screen | Go to MAIN screen to run a test |
| Sample is not attached to the load cell | Attach sample to load cell per drawing | |
| Data measurement consistently low/high | Improper calibration | Check calibration and/or calibration angle |
| Bad calibration | Refer to load cell calibration | |
| Calibration drifts | Bad or damaged load cell | Replace load cell |
| Display is black | Display is bad | Replace display |
| Power switch is not ON | Turn ON power | |
| Power supply is bad | Replace power supply | |
| Display is stuck at ChemInstruments logo | Control board is bad | Replace control board |
Maintenance Procedures
As with all precision instruments, regular care and maintenance are essential to ensure accurate performance and extend the operational lifespan of the equipment. Consistent general cleaning and proper handling not only preserve the integrity of sensitive components but also help prevent malfunctions and minimize the risk of costly downtime. By adhering to recommended maintenance practices, users can maintain reliable test results and enjoy trouble-free operation over the long term.
Touchscreen Cleaning Guidelines
The touch panel is sensitive to chemicals and requires careful cleaning to maintain optimal functionality and appearance. Follow these specific cleaning instructions to ensure safe and effective care:
- Use a soft, lint-free cloth for cleaning. The 3M Microfiber Lens Cleaning Cloth is highly recommended as it effectively cleans the touch panel without the need for liquid cleaners.
- The cloth may be used dry or lightly dampened with a mild cleaner or ethanol. Ensure the cloth is only lightly damp, never wet.
- Never apply cleaner directly to the touch panel surface. If any liquid accidentally contacts the panel, immediately blot it with an absorbent cloth to prevent seepage.
- Use only neutral pH cleaners (neither acidic nor alkaline). Avoid any cleaners containing ammonia, phosphates, or ethylene glycol.
- When applying cleaner, avoid contact with the edges of the film or glass and the flex cable tail to prevent damage.
- Clean the surface gently, following the direction of any surface texture if present.
- Do not use harsh chemicals such as paint thinner, acetone, toluene, xylene, kerosene, or strong alcohols like propyl or isopropyl alcohol in high concentrations.
- Suitable cleaning products are commercially available and specifically formulated for delicate touch panels. Examples include Klear Screen™ and retail brands like Glass Plus® Glass and Surface Cleaner by Reckitt Benckiser.
- Most commercial products contain 1–3% isopropyl alcohol by volume, which is generally safe for use on resistive touch panels.
IMPORTANT: Always check the product's ingredient list carefully. Many cleaners contain unacceptable substances such as ammonia, phosphates, or ethylene glycol, which can damage the touch panel's optical clarity and functionality.